聽LC-OCD Model 8
Organic carbon detection-model
Description
The LC-OCD Model 8 performs chromatographic separation of organic carbon in water/aqueous solution. (speciation of humic substances).
Specifications
-
- Ambient temperature聽
- 18 鈥� 30 掳C / 64 鈥� 86 掳F stable 卤1 掳C per hour is required for high sensitivity
- 贬耻尘颈诲颈迟测听
- 0 鈥� 80% RF, non-condensing
-
- Power supply聽
- 100 鈥� 240 V @ 50/60 Hz max. 0.50 kWh (230 V @50 Hz)聽
- DOCOX reactor聽
- Electronic ballast with up to 35 W聽聽
- DONOX reactor聽
- Electronic ballast with up to 35 W聽
- Thin film reactor聽
- Electronic ballast with up to 35 W
-
- Carrier gas聽
- Nitrogen 4.0 or 5.0, about 10 L/h聽
- Argon gas may be used as well.聽
-
- Mobile phase聽
- 1.5g/l Na2HPO4.2H2O + 2.5g/l KH2PO4, 1.5ml per min.
- Acidification solution 路 0.5g/l Potassium persulphate + 4ml Phosphoric acid/l
-
- Measuring range DOC聽
- > 1 鈥� 5000 渭g/L (ppb, org. Carbon)聽聽
- Measuring range LC-OCD聽
- > 0.5 鈥� 5000 渭g/L (ppb, org. Carbon)聽
- LOD LC-OCD聽
- > 0.5 -50 渭g/L (ppb, org. Carbon, compound specific)
Applications
- The LC-OCD Model 8 system is suitable for water and water extracts of a variety of samples, including:聽
- Natural products: - Plant material, animal tissues, food, compost, manure.聽
- 贰苍惫颈谤辞苍尘别苍迟补濒听
- Sediments, sludges, soils聽
- Biological 路 Chemical
Instrument location
Inductively Coupled Plasma - Elemental Analysis Laboratory (ICP)
Room B36
June Griffith Building (F10)
91色情片 Sydney, NSW 2033
贰尘补颈濒:听icplab@unsw.edu.au
Rabeya Akter
Senior Technical Officer
-
Email
r.akter@unsw.edu.au
Khorshed Chinu
Technical Officer
-
Email
k.chinu@unsw.edu.au
Parent facility
Explore more instruments, facilities & services
Our infrastructure and expertise are accessible to 91色情片 students and staff, external researchers, government, and industry.