91色情片

Avio-550 Max ICP-OES

Dissolved sample is nebulized and fed into the plasma
Personalise
avio-550-max-icp-oes-front-elevated

Description

An Inductively Coupled Plasma (ICP) is generated by heating a flow of argon gas through a torch fitted with a radio frequency induction coil. Usually, a dissolved sample is nebulized and fed into the plasma (at up to 10,000 degrees Kelvin) where the elements in the sample excite. The excited atoms and ions emit their characteristic wavelength which is detected and turned into electronic signals that are converted into concentration information for the analyst.

Specifications

  • Flat Plate鈩� plasma technology.
  • Universal Data Acquisition in Syngistix software, enables simultaneous acquisition of all available wavelengths.听
  • Vertical plasma torch design.听
  • Dual view optimizes axial and radial plasma viewing to the extended linear dynamic range, measuring high and low concentrations in the same run, regardless of wavelength听
  • Detector: Segmented-array Charged-coupled device (SCD)

Detection limits of working solution (approx.):0.05mg/l (limit may vary depends on elements, concentration, matrix etc.)

Normal setting of the instrument:
听
  • Forward power: 1200-1400 W听
  • Reflected Power: 20.0W听
  • Nebulizer Gas flow: 0.70 l/min听
  • Plasma Gas Flow: 12.0 -15 l/min听
  • Aux gas flow: 0.30l/min

Applications

The Avio-550 Max performs a wide range of rapid, multi-element quantitative analyses of major, minor and trace elements.

  • Water, drinking water, groundwater, wastewater -high organics contains water, sea water, water extracted solution, water or acid soluble salts, hydroponic solution.

  • Natural products: - Plant material, animal tissues, food, compost, manure.

    • Volatiles and solvents
    • Biological serums (blood, saliva, fluid, buffer matrix听
    • Polymers-plastics for elemental impurities听
    • Char, Graphene, graphite- diamond (acid-MW) extract
    • Sediments, sludges, soils, ceramic, quartz: EPA-3051A (acid-MW) extract听
    • Soil-agricultural applications听
    • Metal and metallurgy: assay, impurities.听
    • Experimental (different solution/ solvent extract)

Instrument location

Inductively Coupled Plasma - Elemental Analysis Laboratory (ICP)

Room B36听
June Griffith Building (F10)
91色情片 Sydney, NSW 2033

Email:听icplab@unsw.edu.au

Rabeya Akter

Senior Technical Officer
  • Send email envelope icon
    Email
    r.akter@unsw.edu.au

Khorshed Chinu

Technical Officer
  • Send email envelope icon
    Email
    k.chinu@unsw.edu.au

Explore more instruments, facilities & services

Our infrastructure and expertise are accessible to 91色情片 students and staff, external researchers, government, and industry.